Philips Respironics BiPAP Vision Ventilator
Philips Respironics BiPAP Vision Ventilator
Philips Respironics BiPAP Vision Ventilator
The Philips Respironics BiPAP Vision Ventilator is specifically designed to address the unique requirements of Non-Invasive Ventilation within the hospital, including changing leaks, dynamic breathing patterns, and high FIO2 and flow demand.
The BiPAP Vision features accurate monitoring, reliable alarms, and an intuitive menu. Auto-Trak ensures optimum sensitivity despite changing breathing patterns or circuit leaks. Auto-adaptive triggering and cycling reduce work of breathing and support patient-ventilator synchrony, supporting patient compliance and reducing the need for manual adjustments.
The Vision is designed with advanced technology to be accurate, reliable, and easily upgradeable. This low-cost alternate form of sub-acute therapy may reduce the need for intubation for certain patients.
Features
- Internal centrifugal air blower eliminates the need for an air compressor
- Digital Auto-Trak sensitivity automatically adapts to changing leaks on a breath-by-breath basis, allowing the patient’s mask to be fit for comfort rather than a perfect seal
- The integrated display screen displays real-time graphics in waveform or bar scale format to assist with ventilator management
- CPAP and S/T modes are available for maintaining airway patency and providing pressure support ventilation
- Oxygen control allows 21–100% concentrations from without attaching auxiliary oxygen to the mask
- Pressure regulation at the mask ensures accurate therapy for the patient
- The alarm module provides a complete range of alarms that are designed with mask ventilation in mind, minimizing nuisance alarms caused by mask leak
- Small size allows for easy counter top operation or use with rolling stand
- Includes rolling stand and support arms
- One (1) Year Warranty
Dimensions At The Base
- Height: 10.625 in (27 cm)
- Width: 10.375 in (36.5 cm)
- Length: 16 in (40.6 cm)
- Weight: 34 lbs (15.4 kg)
Specifications
- Ventilation Modes: CPAP, S/T, PAV/T
- Parameters: IPAP, EPAP, CPAP, Rate, Timed Inspiration, Rise Time, FiO2
- Alarms: High/Low Pressure, Apnea, Low Minute Ventilation, High/Low Rate
- Display: IPAP, EPAP, CPAP, Rate, VT, Minute Ventilation, Total Leak, Patient Lead, PIP, % Triggered Breathes, Ti/Ttot
Electrical
- AC Input Voltage (VAC):
- 100/120/230/ 240 VAC
- Single Phase
- ±10%
- Fuses:
- 100 – 120 VAC ~ T 3.5 A, 5 × 20 mm, Time Lag (×2) (For serial no.’s 100500 and higher – Respironics Reorder # 1000749)
- 115 VAC ~ T 3.0 A, 250 V, ¼” × 1¼” (For serial no.’s 100499 and lower – Respironics Recorder # 582100)
- 220 VAC, 230 VAC and 240 VAC ~ T 1.6 A, 250 V, 5 × 20 mm (For all serial no.’s – Respironics Recorder # 1000750)
- Power Consumption: 300 VA max.
- AC Current: 3.0 A maximum
- AC Frequency: 50/60 Hz
Modes
CPAP
- Control Range: 4 to 20 cm H2O
- Control Increments: 1 cm H2O
S/T
- Control Range: 4 to 40 cm H2O
- Control Increments: 1 cm H2O
PAV/T
- Control Range: Quick Start Menu
- Control Increments: Obstructive, Restrictive, Mixed, Normal
Control Ranges & Parameters
IPAP
- Control Range: 4 to 40 cm H2O
- Control Increments: 1 cm H2O
EPAP
- Control Range: 4 to 40 cm H2O
- Control Increments: 1 cm H2O
CPAP
- Control Range: 4 to 40 cm H2O
- Control Increments: 1 cm H2O
Rate
- Control Range: 4 to 40 BPM
- Control Increments: 1 BPM
Timed Inspiration
- Control Range: 0.5 to 3.0 sec
- Control Increments: 0.1 sec
IPAP Rise Time
- Control Range: 0.05 to 0.40 sec
- Control Increments: 4 set points: .05, .1, .2, .4 sec
Oxygen Concentration (%O2)
- Control Range: 21 to 100%
- Control Increments: 4% from 21 to 25% 5% from 25 to 100%
Exhaled Tidal Volume
- Display Range: 0 to 4000 ml
- Display Resolution: 1 ml
Minute Ventilation
- Display Range: 0 to 99 L/min
- Display Resolution: 1 L/min
Total Leak
- Display Range: 0 to 300 L/min
- Display Resolution: 1 L/min
Patient Leak
- Display Range: 0 to 300 L/min
- Display Resolution: 1 L/min
Peak Inspiratory Pressure (PIP)
- Display Range: 0 to 50 cm H2O
- Display Resolution: 1 cm H2O
% Of Patient Triggered Breaths
- Display Range: 0 to 100%
- Display Resolution: 1%
Ti/Ttot
- Display Range: 0 to 100%
- Display Resolution: 1%
Alarms
High Pressure
- Control Range: 5 to 50 cm H2O
- Control Increments: 1 cm H2O
Low Pressure
- Control Range: Disabled; 0 to 40 cm H2O
- Control Increments: 1 cm H2O
Apnea
- Control Range: Disabled; 20 to 40 sec
- Control Increments: 4 set points: Disabled, 20, 30, 40 sec
Low Minute Ventilation
- Control Range: Disabled; 1 to 99 L/min
- Control Increments: 1 L/min
Respiratory Rate
- Control Range: 4 to 120 BPM
- Control Increments: 4 set points: 1 BPM